What is the calibration parameters of HPLC.
Calibration parameters of HPLC are used to ensure accurate and precise measurements of analytes.
Calibration parameters include retention time, peak area, and peak height.
Retention time is the time it ...read more
There are a total 9 parameters for HPLC calibration, these are as follows;
Flow Rate Accuracy
Injector Precision
Injector Linearity
Wavelength accuracy of the detector
Detector Linearity
Gradient Composition Accuracy
Carry Over
Temperature Accuracy
Drift and Noise
Calibration of HPLC is performed on six month basis or after any major failure or after maintenance. Acceptance Criteria - Wavelength maxima found should be between 273 ± 2 nm. Wavelength maxima found...read more
Flow rate: 1.0 ml/min. Wavelength For UV: 254 NM. Wavelength For PDA: 254 nm (2D data collection at 3.6 nm bandwidth) Injection volume: 0.0 ml.
Flow rate: 1.0 ml/min. Wavelength For UV: 254 NM. Wavelength For PDA: 254 nm (2D data collection at 3.6 nm bandwidth) Injection volume: 0.0 ml.
Flow rate: 1.0 ml/min. Wavelength For UV: 254 NM. Wavelength For PDA: 254 nm (2D data collection at 3.6 nm bandwidth) Injection volume: 0.0 ml.